Electro-polymerization of poly Eugenol on Ti and Ti alloy dental implant treatment by micro arc oxidation using as Anti-corrosion and Anti-microbial

 

Haider Abdulkareem AlMashhadani1, Khulood Abed Saleh2

1Al Rasheed University College, Dentistry Department

2Department of Chem., College of Science, Baghdad University, Jaderyah, Baghdad, Iraq

*Corresponding Author E-mail: H_R200690@yahoo.com

 

ABSTRACT:

In this work, electrochemical process was presented to polymerized eugenol on Gr.2 and Gr.5 titanium alloys before and after treated by Micro Arc Oxidation (MAO), where Gr.2 is commercial pure titanium and Gr.5 is Ti-6Al-4V dental alloys. The deposited layers were characterized by scanning electron microscopy (SEM), energy-dispersive X-ray spectroscopy (EDS), X-ray diffraction (XRD), and Fourier transform infrared spectroscopy (FTIR). The adhesion strength of polymeric thin-film was estimation by using pull-off adhesion test and the result was the adhesion strength of PE was (1.23 MPa) on Gr.2 before MAO and increase to (1.98 MPa) on Gr.2 after MAO treatment. The corrosion behavior of Gr.2 and Gr.5 alloy in artificial saliva environment at temperature ranged 293-323K has been studied and assessed by means of electrochemical technique spotentiodynamic polarization curves. The corrosion protection increased to 63% and 80% after coated blank Gr.2 and Gr.5 by PE, while when deposition PE on treated Gr.2 and Gr.5 by MAO, the protection efficiency (I%) increase to 68% and 81% respectively at 323K. The antimicrobial activities of the samples were evaluated against different bacteria and oral fungi (Candia), PE coating show antibacterial activity (28 and 25 mm) against S. aureus and B. Subtilis respectively and show antifungal activity (29 and 24 mm) against C. albicans and C. glabrata oral fungi respectively.

 

KEYWORDS: PolyEugenol, Eugenol, Titanium alloy, MAO.

 

 


1.    INTRODUCTION:

Micro arc oxidation (MAO) is one of the most promising techniques for metal and alloy surface therapy and has subsequently been widely accepted by multiple industries branches. MAO is usually used on valve materials such as Al, Mg, Ti, Ta, Nb, Zr, and Be to create multi-purpose wear, corrosion, and heat-resistant dielectric and ornamental coatings(1-4). In recent years, titanium and Ti-based alloys are widely used in the field of biomedical materials, such as dental restorations, orthopaedic prosthetics, dental implant and heart valves (5-8) because of their good biocompatibility and good corrosion resistance properties deriving from the formation of oxide films on surface of metals.

 

It is unimpeachable that the degradation of Ti and its alloys will occur by electrochemical reactions at the interface between surface of titanium alloys and saliva, inducing in the breakdown of protective titanium oxide passive films, the release of titanium ion and the decrease of the service life of titanium alloys. In recent years, many ways based on a surface modification to improving the biocompatibility of Ti-based implants have been suggested(9). MAO's use favours the adaptation of surface composition, crystallographic structure, and morphology to achieve large features that parent metal cannot provide. The MAO process mechanism is based on the anodizing electrochemical reaction that happens on a metallic surface and is followed by micro-arc discharge to create an oxide ceramic surface layer with specific morphology and phase composition(10,11). Also, coating metals with polymer films is a useful strategy that is of great interest for modification of Ti and its alloys in dental applications since it can modify the properties of the metal surface and/or protect it from corrosion. Electrochemical polymerization is the best technique for manufacturing thin films on metal surfaces in order to protect them from corrosion since it enables the synthesis of polymeric films directly onto the metal surface starting from solutions of the relevant      monomers (12-14).

 

The significant element of clove oil is eugenol (4-allyl-2-methoxyphenol), a methoxyphenol with a brief hydrocarbon chain(15). Eugenol has been used as a spice because of its strong odour and as a dental antiseptic. It has been widely used for sedation in patients with toothache, pulpitis, and dental hyperalgesia as a therapeutic agent in dentistry(16). Pharmacologic studies have demonstrated that eugenol has local anaesthetic(17), bacteriostatic and bactericidal,(18) anticandidal,(19) and antifungal(20) properties.

 

In this work, the ceramic coating is formed on pure titanium (Gr.2) and Ti-6Al-4V alloys (Gr.5) in alkaline electrolyte then forms a thin film of poly eugenol (PE) on modification Ti and alloys surface by electro-polymerized of eugenol. The electrosynthesis of polyeugenol (PE) coatings were characterized by FTIR, X-ray and scanning electron microscopy (SEM), coating adhesive where measured in order to assess their surface morphologies. The biological activity for PE coating against several bacteria and oral fungi were studied also.

 

2.    MATERIALS AND METHODS:

2.1. Titanium Preparation:

Titanium specimens of 2 ×2 cm2 area were obtained from commercially pure titanium sheet (Gr.2) of 0.5mm thickness, while circular shape samples of Ti-6Al-4V alloys (Gr.5) were mechanically cut with dimensions of 1cm diameter and 4-5 mm thickness. These samples were polishes with emery papers in different grade include 600, 800, 1200 and 2000 mesh grit, and then washed with tap water, distilled water then ethanol and finally with acetone then dried by using a hair drier. Before electrochemical polymerization, titanium specimens were treatment by heating (500oC) to form a barer titanium oxide thin film which increase the current between electrode and eugenol solution.

 

2.2. Artificial Saliva Preparation:

Reference electrolyte was used named; modified Fusayama artificial saliva which prepared according to a method that has been described in a previous detail report(21), which closely resembles natural saliva, with composition of (0.4g/l KCl, 0.4g/l NaCl, 0.906g/l CaCl2.2H2O, 0.69g/l NaH2PO4.2H2O, 0.005g/l Na2S.9H2O and 1g/l urea)., and pH of this electrolyte was 6.2.

 

2.3. Micro Arc Oxidation of Ti and Ti alloys:

For MAO an AC power of 150 V was employed for (180 sec.). A titanium plate was used as an anode with a large titanium sheet as the cathode electrode in electrolyte contained 10g/L KH2PO4 and 0.5g/L KOH, The electrolyte temperature was kept constant at 25oC. After the MAO process, the samples were washed with distilled water and dried at room temperature. Figure (1) shows MAO cell.

 

Figure 1. Micro Arc Oxidation cell.(22)

 

2.4. Electrochemical Polymerization of Eugenol:

After heating titanium electrode, it was placed in 0.1M NaOH containing 10mM eugenol and it is necessary to use a large sheet of titanium as counter electrode, the voltage of (3 V) was applied at room temperature with duration (90 min.), then the electrode washed by distilled water and dry by hair drier.

 

2.5. Characterization and test:

Spectroscopic analysis of eugenol and polyeugenol coated formed with Fourier transform infrared spectroscopy. The surface morphologies of PE coating were measured and analyzed by XRD and SEM. The galvanostatic polarization was measured using WENKING M Lab. (Germany) device. Polarization curves were achieved for dental titanium alloys before and after MAO treatment. TGA of the obtained coated polymers was performed using a Shimadzu DT-30 thermal analyzer (Shimadzu, Kyoto, Japan). And the weight loss for the polymeric coating (PE) was measured at temperature ranging from 25 up to 600oC, at the rate of 10°C/min to determine the degradation rate of the polymer film. The PE coating was tested for antimicrobial activity by well diffusion method against pathogenic bacteria such as Gram-negative [Escherichiacoli (E-coli), Acinetobacter baumannii (A-baumannii)] and Gram-Positive [Staphylococcus aureus (S-aureus), and Bacillus subtilis (B-subtilis)]. and coating was tested for antimicrobial activity by well diffusion method against and oral fungi such as Candida albicans (C-Albicans), Candida glabrata (C-glabrta), Candida parapsilosis (C-parapsiosis) and Candida tropicalis (C-tropicalis).

 

3.    RESULT AND DISCUSSION:

3.1. Chemical Structure and Morphology of the Polymeric Eugenol:

Monomeric eugenol exists in the form of a corresponding phenolate anion under conditions applied in this work (0.1 M NaOH). It is well documented that phenolate ions undergo electrochemical oxidation to phenolate radicals through one-electron oxidation(23). These radicals are known to be produced from unsubstituted phenol, resulting primarily in para-coupling, resulting in linear polyphenylene oxide(24). On the other hand, two-electron oxidation accompanied by the loss of the methoxy-substituent would produce 4-allyl-1,2-quinone(25). As known from the literature 1,2-quinones are extremely unstable species and undergo a variety of intermolecular reactions leading to an ill-defined highly cross-linked polymeric structure(25). Probably, the allyl group is also involved in the polymerization since it is expected to be susceptible to the attack of free radicals and to take part in so-called Diels-Alder reactions between 1,2- quinone (diene) and the allyl C=C bound (dienophile)(26). Finally it should be mentioned that at sufficiently high electrodic potential an oxidative degradation of the polymer may occur which will result in the creation of additional oxygen-containing functionalities most likely carbonylic and carboxylic ones or in an extreme case in the generation of carbon dioxide whose presence in the polymeric structure is clearly seen by infrared spectroscopy as discussed below.

 

To get more information about the structure of the PE formed on titanium plate and to identify functional groups present in it, reflectance FT-IR spectra were recorded for the monomer and polymer formed on a titanium plate by electrochemical polymerization, which is shown in Figure (2).


 

Figure 2. FTIR spectra of a) Eugenol, b) Hypothesized PolyEugenol.

 


Bands at 3465 and 3436 cm-1, respectively, are assigned to the hydroxyl stretch groups in phenol, and the band at 1539 cm-1 possibly indicate for C=O stretch of sodium carboxylate salts of the polymer in alkaline solution. The band at 1639 cm-1 may be assigned to the C=O stretch of carbonate groups included in the polymeric film on titanium plate.

 

Examination of the electropolymerized PE films on Gr.2and Gr.5 surfaces before MAO treatment using scanning electron microscopy (SEM) is shown in Figure (3). As seen in the photograph the PE thin film was accumulation on titanium surface before MAO. After magnification to 10K and 100K X shaped structure of PE is evenly distributed hexagonal Nano-structures that tend to aggregate in characteristic cluster forms with intervening porosity. The average partials size of PE deposited on Gr.2 and Gr.5 about (32nm) and (49nm) respectively. Figure (3eandf) show the cross-section SEM for the thin film of PE and enable measurements of the thickness of poly thin-film to reach to (445nm) and (282nm) for Gr.2 and Gr.5 respectively.


 

Figure 3. Scanning electron micrographs of PE films deposited on (a and b) Gr.2 (c and d) Gr. 5 and Cross Section Scanning electron micrographs of PE films deposited on (e) Gr.2 (f) Gr.5.

 

Figures (4) show Gr.2 and Gr.5 alloys after MAO treatment with and without PE films on their surfaces, Figure (4) shows the deposition of PE on the surface of titanium after 150 V MAO process, from the cross-section (Fig. 6d), the PE deposition in the porous Gr.5 surface. The average partials size of PE deposited on Gr.2 and Gr.5 after MAO treatment around (33nm) and (58nm) respectively.

 

Figure 4. Scanning electron micrographs of PE films deposited on a, b) Gr.2 after MAO treatmentc, d) Gr.5 after MAO treatment.

 


The X-ray diffraction features of the blank Gr.2 and for PE formed on Gr.2 by electro-polymerization method are displayed in Figure (5), the XRD pattern for the Gr.2 coated by PE before and after MAO is shown in Figure (5B and 5C), and the peak positions are slightly shifted to lower 2-theta (2θ) values compared to that uncoated Gr.2 and The two high-intensity peaks in Figure (5B) located at 2θ are (25.1o and 25.3o) of polymeric film of eugenol on titanium plate but a small peak in Figure (5C) located at 2θ are (25.38o) may be for PE or formation of titanium oxide layer from MAO process. The crystallite size of the polymeric film was calculated from the full width at half maximum (FWHM) of the peaks using Debye–Scherrer’s approximation (Eq. 1)

 

 

Where d is the crystallite size, k is the wavelength of CuKa radiation (k = 1.542A°), β is FWHM for the diffraction peak under consideration (in radians), θ is the diffraction angle, and k is the broadening constant (k=0.9).

 

Figure 5.XRD patterns for (a) Gr.2, (b) Gr.5.

 


The crystallite size has been detected for pure titanium and have the average size equal to (227.5nm), where the crystalline size for PE peak ((25.1o and 25.3o) was equal to (18.99nm) while the size of the average partial of P.E deposited on pure Ti in SEM was about (32 nm).

 

3.2. Corrosion test:

The Tafel curves for the corrosion test for pure titanium (Gr.2) and for (Gr.5) immersed in artificial saliva before and after electrochemical polymerization by PE at different temperatures were recorded and plotted in Figure (6). The polarization curves of uncoated (Gr.2) show corrosion potential ranged between (-146.7 to -165.7 mV) and this potential shifted to (-743.2mV) at 293K after coated (Gr.2) by PE, while Ecorr of uncoated (Gr.5) alloy were ranged between (-556.3 to -485.5mV) then these potentials shifted to more noble potential when Gr.2 surface coated by PE were ranged between (-95 to -149.9mV) Figure (6). Ecorr for Gr.5 at temperature range (293-323K) have high negative values compared with Ecorr values for Gr.2, coating Gr.5 led to increase Ecorr values to noble direction this indicate that Ti alloys are difficult to corrode after coated by PE. It is clear in figure (6b) that Ecorr for Gr.2 coated by PE were shifter to more noble direction with increasing temperature from 293 to 323 K.


 

Figure 6. Polarization curves for a) Uncoated Gr.2, b) Gr.2 Coated by PE, c) Uncoated Gr.5 and d) Gr.5 Coated by PE.

 


Data in the table (1) shows that the corrosion current density, slightly increased with temperature increased for both Gr.2 and Gr.5 but after coated by PE the corrosion current density for Gr.2 have not any effect with increasing temperature, while the icorr for Gr.5 was increased slightly with increased temperature which were ranged between (0.6 to 0.679µA.cm-2), this results indicate that the efficiency of PE against corrosion increased at high temperature.

 

The Protection Efficiency (%I) can be calculated by using the equation(26, 27):

 

 

Where (icorr)b and (icorr)p are the corrosion current density (μA.cm-2) for Blank Gr.2 and Gr.5 and for protected Gr.2 and Gr.5 respectively.

 

The best %I were obtained after coated Gr.2 by PE which reached to 50% at 293K, while the best %I which were obtained after coated Gr.5 by PE ranged between (70-80) % table (1) data shows that %I slightly increased with increasing temperature, that indicate icorr for Ti coated by PE was not affected by increasing temperature while it lead to increase efficiency of polymer deposit on titanium against corrosion.

 

Ecorr values for Gr.2 were higher than Ecorr values for Gr.5, while Ecorr values for Gr.5 coated by PE were higher than Gr.2 coated by PE, and the %I obtained by PE coating for Gr.5 higher than %I for PE coating for Gr.2.

 

Coating by PEled to a protective layer against temperature for the two samples Gr.2 and Gr.5 where the %I increase with temperature increased, and the highest %I value is 80.37% for coated Gr.5 by PE at 323K. Polarization near the corrosion potential Ecorr is conducted to determinate the polarization resistance on the basis of the following equation, i.e., the Stern–Geary equation:

 

 

where Rp is the polarization resistance of the system, ΔE is the difference between the polarization potential E and the corrosion potential Ecorr, Δi is the difference between the measured current density i and the corrosion current density icorr, and ba and bc are the anodic and cathodic Tafel coefficients, respectively.

 

Polarization resistance has such as requirements to be discussed and the measurements of full polarization curves are particularly helpful for identifying corrosion trouble and initiates reconditioned action(28,29).

 

Figure (7) depicts the polarization curves for the corrosion of samples after MAO treatment with and without coated by P.E. The fitted values of the corrosion kinetics parameters for all sample are listed in the Table (2).

 

The data shows the Ecorr of Gr.2 after MAO treatment ranged between (-388.6 to -294.1mV) while after coated by P.E the Ecorr shifted to a more noble direction (95 to 17.4 mV). Ecorr of Gr.5 after MOA treatment ranged between (-388.2 to -343.7mV) also shifted to more noble direction when coated by P.E (109.2 to 186.7mV). Figure (7) show the corrosion potential of Gr.2 shifted to more noble direction with temperature increasing except Gr.2 coated by PE after MAO treatment shifted to more active potential.


 

Table 1. Corrosion kinetic parameters for Gr.2 and Gr.5 before MAO in artificial saliva at different temperature in the range (293-323)K.

Temp./K

-Ecorr/ mV

Icorr/µA.cm-2

-ba/ mV.dec-1

bc/ mV.dec-1

%I

Rp/Ω.cm2

Blank of Gr.2

293

-146.7

1.41

226

336.4

-

41629.96

303

-151.0

1.50

258.1

423.3

-

46413.99

313

-165.7

1.60

282.8

455.5

-

47350.11

323

-147.7

1.65

249

574.8

-

45721.07

Gr.2coated

by PE

293

-743.2

0.702

70.3

204.0

50.21

32339.153

303

-688.4

0.669

80.9

122.8

55.40

31654.52

313

-598.9

0.667

99.2

188.2

58.31

42288.74

323

-555.8

0.602

102.7

171.8

63.52

46361.85

Blank of Gr.5

293

-556.3

2.010

144.8

301.5

-

21131.92

303

-572.8

2.590

149.8

368.4

-

17854.19

313

-577.8

3.010

205.8

489.3

-

20898.39

323

-485.5

3.460

241.3

532.8

-

20842.72

Gr.5coated

by PE

293

-95.4

0.600

18.1

17.7

70.14

6476.25

303

-75.4

0.636

16.3

15.6

75.44

5442.15

313

-120.2

0.650

15.4

18.0

78.40

5544.21

323

-149.9

0.679

19.3

18.0

80.37

5956.03

 


Table (2) show corrosion kinetics parameters for the two titanium samples treated by MAO coated and uncoated by PE. icorr values for Gr.2 after MAO treatment were ranged (0.692 – 1.08)µA.cm-2 while after coated by PE the icorr reduced to (0.495 – 0.519)µA.cm-2 at temperature ranged (293-323)K. icorr values for Gr.5 after MAO treatment were higher than Gr.2, ranged (1.17 – 1.93)µA.cm-2 while after coated by PE decreased to (0.569 – 0.644) µA.cm-2, this reduces in icorr which give %I reach to 64% and 71% for Gr.2 and Gr.5 respectively and this value increases with temperature increased reach to 68% and 81% for Gr.3 and Gr.5 respectively, where icorr slightly effected by temperature.


 

Figure 7. Polarization curves for a)Gr.2 after MAO, b) Gr.2 Coated by PE after MAO, c)Gr.5 after MAO and d) Gr.5 coated by PE after MAO.

 

Table 2. Corrosion kinetic parameters for Gr.2 and Gr.5 after MAO treatment in artificial saliva at different temperature in the range (293-323)K.

Temp./K

Ecorr/ mV

Icorr/µA.cm-2

-ba/ mV .dec-1

bc/ mV .dec-1

%I

Rp/Ω.cm2

Gr.2 after MAO

293

-388.6

0.692

194.9

335.8

50.92

77382.61

303

-471.0

0.841

136.7

249.8

43.93

45616.45

313

-344.0

0.914

147.8

239.5

42.87

43420.25

323

-294.1

1.080

152.9

294.0

34.54

40441.45

Gr.2coated

by PE after MAO

293

95.0

0.495

98.9

110.9

64.89

45858.85

303

35.4

0.504

66.3

99.7

66.40

34306.47

313

25.2

0.511

61.8

74.9

68.06

28773.11

323

17.4

0.519

51.3

67.4

68.54

24370.54

Gr.5 after MAO

293

-382.2

1.170

19.2

36.7

41.79

4678.16

303

-345.3

1.290

9.8

26.4

50.19

2405.68

313

-347.9

1.710

22.7

21.7

43.18

2817.17

323

-343.7

1.930

26.2

54.0

44.21

3968.89

Gr.5coated

by PE after MAO

293

109.2

0.569

121.1

153.9

71.69

51718.25

303

118.7

0.588

133.6

124.6

77.29

47609.87

313

184.8

0.629

108.7

105.9

79.10

37029.78

323

186.7

0.644

100.2

99.3

81.38

33627.54

 


3.3. Kinetic parameters for the corrosion process:

From logarithm of corrosion current density plotted against reciprocals temperature for the corrosion of uncoated, coated Gr.2 and Gr.5 by PE before and after MAO treatment can obtain the data in the table (3) and shows the activation energy increased after MAO treatment, change from (4.232 kJ.mol-1) to (11.173 kJ.mol-1) and from (14.056 kJ.mol-1) to (15.833 kJ.mol-1) for Gr.2 and Gr.5 respectively, while after coated titanium by P.E before and after MAO treatment, the activation energy decreased and Arrhenius factor also decrease that indicated decrease in the number of corrosion sites on Gr.2 and Gr.5 alloys surface.

 

Table 3. Kinetic parameters for Gr.2 and Gr.5 in artificial saliva.

 

Gr.2

Gr.5

 

Ea/kJ.mol-1

A x1024 Molecules.cm-2 .S-1

Ea/kJ.mol-1

A x1026 Molecules.cm-2 .S-1

Blank

4.232

4.84

14.056

3.98

Post MAO

11.173

41.3

15.833

3.21

Coated by PE

3.622

0.096

2.792

0.016

Coated by PE post-MAO

1.277

0.493

2.792

0.016

 


3.4. The Thermodynamic Studies:

The change in the entropy (ΔS*) for transition state of corrosion process could be derived according to modified Arrhenius equation:

 

 

Where icorr is the corrosion current density of titanium in artificial saliva derived from Tafel plot, h is the Plank’s constant, N is the Avogadro number, ΔS* is the activation entropy and ΔH* is activation enthalpy. The plot of log icorr/T against 1/T obtained straight lines with the slope and an intercept.

 

Values of ΔS* which presented in the table (4) reflect the changes in the order and transition state orientation of corrosion process of Gr.2 and Gr.5, and the data in the Table (4) shows the values of ΔS* were slightly effected by protected coating. Enthalpy of activation ΔH* is a component of activation energy, for this note the values of ΔH* linked to the values of (Ea). The coated Gr.2 by PE before and after MAO treatment has ΔH* changed from endothermic to exothermic, while coated Gr.5 by PE before and after MAO led to decrease the value of activation enthalpy, but the treatment of Gr.2 and Gr.5 by MAO cause an increase in ΔH*values.

 

The free energy change of electrochemical corrosion reaction transition state can be calculated using the following equation:

 

The values of ΔG* took positive values and showed almost small change with increasing temperature, indicating that the activated complex was not stable and the probability of its formation decreased somewhat with rising in temperature(30)


 

Table 4. Kinetic and thermodynamic parameters for Gr.2 and Gr.5 in artificial saliva at temperature range (293-323)K.

 

 

ΔG*/kJ.mol-1

ΔH*/

kJ.mol-1

ΔS*/

kJ.mol-1 .K-1

293

303

313

323

Gr.2

Blank

70.87

73.23

75.59

77.96

1.675

-0.236

Post MAO

72.59

74.77

76.95

79.14

8.616

-0.218

Coated by P.E

72.56

75.25

77.94

80.63

-6.178

-0.268

Coated by P.E post-MAO

73.43

75.98

78.53

81.09

-1.329

-0.255

Gr.5

Blank

69.96

71.95

73.95

75.94

11.499

-0.199

Post MAO

71.41

73.46

75.50

77.55

11.456

-0.205

Coated by P.E

73.22

75.70

78.18

80.66

0.541

-0.248

Coated by P.E post-MAO

72.91

75.39

77.87

80.35

0.236

-0.248

 


3.5. TG/DSC analysis:

TG /DSC curves for PE thin film formed on titanium show weight loss at four regions as shown in Figure (8). The first one occurs in the region between 20 and 75oC which is assigned to the loss of weight of about 5% due to absorbed water. The second stage recorded a mass change of 10% at 186oC, and is the loss of water of lattice accompanied by a small endothermic at peak 299 oC. The gradual loss of mass from 300 to 600oC is due to the liberation of nitrates and carbonates from precursors. The total mass loss is approximately 42%.

 

Figure 8. TG/DSC curves for poly eugenol coated on Ti.

 

3.6. Adhesion Test:

Excellent adhesion of a polymeric coated film on the Ti surface led to good corrosion protection of Ti and Ti alloy. Adhesion testing determines how well a coating is bonded to the Ti surface. The adhesion pull-off testing was performed for different coating types on commercially pure titanium (Gr.2) only because the test is carried out on samples with of dimensions 4 x 4 cm. The values for the failure strength and the type of failure indicate that the PE coating after MAO treatment reach to (1.98 MPa) has stronger adhesive properties than P.E coating on Ti surface before MAO treatment (1.23 MPa) by approximately 0.75 MPa.

 

3.7. Antimicrobial Activity:

With the expanding rate of microbial species in the oral cavity becoming resistant to present antibiotics, many studies have attempted and effective antimicrobial reagents of resistance and with low cost. Such problems have led to the resurgence in the use of metal-based antiseptics which are applicable to broader spectrum of bacteria with far lower susceptibility to induce resistance than common antibiotics. The positive control amoxicillin was highly effective against positive gram bacteria and lower effective against negative gram bacteria at 1 mg/mL concentration of and the negative control DMSO does not have any activity against both gram-positive and negative bacteria. The highest antibacterial activity was observed against B. subtilis for PE coating followed against B. subtilis and PE coating against S. aureus. The antibacterial activity of PE coating against E. coli was higher than activity of positive control, as shown in table (5).


 

Table 5. Antibacterial activity (zone of inhibition, mm) of the PE coating.

Bactria

E. coli

A. baumannii

B. subtilis

S. aureus

Amoxicillin (+ve)

13

14

31

25

PolyEugenol

19

13

25

28

Fungi

C. albicans

C. tropicalis

C. glabrata

C. parapsilosis

Fluconazole (+ve)

28

22

22

22

PolyEugenol

29

20

24

20

 


The positive control Fluconazole was highly effective against Candida species at the concentration of 1 mg/ml. The antifungal activity of coating was tested against four Candida species (C. albicans, C. tropicalis, C. glabrata and, C. parapsilosis) by well diffusion method. The highest zone of inhibition was observed in PE coating against C. albicans, as shown in table (5).

 

CONCLUSION:

Electro-polymerization of eugenol on Gr.2 and Gr.5 surfaces and treatment the surfaces of Gr.2 and Gr.5 by MAO technique were employed to increase the corrosion resistance of these alloys in artificial saliva. From the obtained results, the following conclusions are obtained:

1.     The particles size of PE on Gr.2 before and after MAO was around (33nm), not affected by MAO process, while the particles size of PE on Gr.5 increase from (49nm) to (58nm) after treated by MAO.

2.     Coating Gr.5 by PE led to shifted the corrosion potential to more noble potential, while Gr.2 coated by PE shifted to more active potential and this coating led to reduce the corrosion current density to 50% and 70% at 293K for Gr.2 and Gr.5 respectively, and it have not any effect with increasing temperature. The protection efficiency of PE against corrosion reaches to 63% and 80% at 323K respectively.

3.     Coating Gr.2 by PE after MAO treatment led to increase the protection efficiency to 65% at 293K and reach to 69% at 323K, while Gr.5 coated by P.E after MAO have not largely effect on %I.

4.     The activation energies for the corrosion of treated alloys by MAO were increased, but coated alloys by P.E pre and post-MAO led to decrease A and Ea values, the decreased of A values to decrease in corrosion active sites number on the surface of alloys coated by PE.

5.     The adhesion strength of PE film on Gr.2 increased when coated on a treated surface by MAO.

6.     The antibacterial activity of PE coating was highest against positive gram bacteria (S. aureus and B. subtilis) than a negative gram (E. coli and A. baumannii).

7.     The antifungal activity of PE coating shows good activity against oral fungi (candida species).

 

REFERENCES:

1.      Santos-Coquillat A, Tenorio RG, Mohedano M, et al. Tailoring of antibacterial and osteogenic properties of Ti6Al4V by plasma electrolytic oxidation. Applied Surface Science. 2018; 454:157-72.

2.      Fei C, Hai Z, Chen C, et al. Study on the tribological performance of ceramic coatings on titanium alloy surfaces obtained through microarc oxidation. Progress in Organic Coatings. 2009; 64(2-3):264-7.

3.      Wang Y, Jiang B, Lei T, et al. Dependence of growth features of microarc oxidation coatings of titanium alloy on control modes of alternate pulse. Materials Letters. 2004; 58(12-13):1907-11.

4.      Yerokhin A, Nie X, Leyland A, et al. Characterisation of oxide films produced by plasma electrolytic oxidation of a Ti–6Al–4V alloy. Surface and Coatings Technology. 2000; 130(2-3):195-206.

5.      Albayrak Ç, Hacısalihoğlu İ, Alsaran A. Tribocorrosion behavior of duplex treated pure titanium in simulated body fluid. Wear. 2013; 302(1-2):1642-8.

6.      Rautray TR, Narayanan R, Kim K-H. Ion implantation of titanium based biomaterials. Progress in Materials Science. 2011; 56(8):1137-77.

7.      Kumar S, Narayanan TS. Electrochemical characterization of β-Ti alloy in Ringer's solution for implant application. Journal of alloys and compounds. 2009; 479(1-2):699-703.

8.      De Almeida L, Bastos I, Santos I, et al. Corrosion resistance of aged Ti–Mo–Nb alloys for biomedical applications. Journal of Alloys and Compounds. 2014; 615: S666-S9.

9.      Liu X, Chu PK, Ding C. Surface modification of titanium, titanium alloys, and related materials for biomedical applications. Materials Science and Engineering: R: Reports. 2004; 47(3-4):49-121.

10.   Yerokhin AL, Lyubimov VV, Ashitkov RV. Phase formation in ceramic coatings during plasma electrolytic oxidation of aluminium alloys. Ceramics International. 1998; 24(1):1-6.

11.   Yerokhin A, Nie X, Leyland A, et al. Plasma electrolysis for surface engineering. Surface and coatings technology. 1999; 122(2-3):73-93.

12.   AlMashhadani HA, editor. Corrosion Protection of Pure Titanium Implant by Electrochemical Deposition of Hydroxyapatite Post-Anodizing. IOP Conference Series: Materials Science and Engineering; 2019: IOP Publishing.

13.   Mandracci P, Mussano F, Rivolo P, et al. Surface treatments and functional coatings for biocompatibility improvement and bacterial adhesion reduction in dental implantology. Coatings. 2016; 6(1):7.

14.   De Giglio E, Cometa S, Cioffi N, et al. Analytical investigations of poly (acrylic acid) coatings electrodeposited on titanium-based implants: a versatile approach to biocompatibility enhancement. Analytical and bioanalytical chemistry. 2007; 389(7-8):2055-63.

15.   Szabadics J, Erdelyi L. Pre-and postsynaptic effects of eugenol and related compounds on Helix pomatia L. neurons. Acta Biologica Hungarica. 2000; 51(2-4):265-73.

16.   Francis LE, Wood DR. Dental pharmacology and therapeutics: Saunders; 1961.

17.   Brodin P, Røed A. Effects of eugenol on rat phrenic nerve and phrenic nerve-diaphragm preparations. Archives of Oral Biology. 1984; 29(8):611-5.

18.   Walsh SE, Maillard JY, Russell A, et al. Activity and mechanisms of action of selected biocidal agents on Grampositive andnegative bacteria. Journal of applied microbiology. 2003; 94(2):240-7.

19.   Chami N, Bennis S, Chami F, et al. Study of anticandidal activity of carvacrol and eugenol in vitro and in vivo. Oral microbiology and immunology. 2005; 20(2):106-11.

20.   Lee S-J, Han J-I, Lee G-S, et al. Antifungal effect of eugenol and nerolidol against Microsporum gypseum in a guinea pig model. Biological and Pharmaceutical Bulletin. 2007; 30(1):184-8.

21.   Al Mashhadani HAY. Study the Effect of Punica Granatum as Oral Antifungal on the Corrosion Inhibition of Dental Amalgam Alloy in Saliva. 2018.

22.   Ibrahim MZ, Sarhan AA, Yusuf F, et al. Biomedical materials and techniques to improve the tribological, mechanical and biomedical properties of orthopedic implants–A review article. Journal of Alloys and Compounds. 2017; 714:636-67.

23.   Kyriacou DK. Modern electroorganic chemistry: Springer Verlag; 1994.

24.   Bartlett P, Cooper J. A review of the immobilization of enzymes in electropolymerized films. Journal of Electroanalytical Chemistry. 1993; 362(1-2):1-12.

25.   Ueda C, Tse DC-S, Kuwana T. Stability of catechol modified carbon electrodes for electrocatalysis of dihydronicotinamide adenine dinucleotide and ascorbic acid. Analytical Chemistry. 1982; 54(6):850-6.

26.   Abbas HA, Alsaade KAS, AlM Ashhdan HAY, editors. Study the Effect of Cyperus rotundus Extracted as Mouthwash on the Corrosion of Dental Amalgam. IOP Conference Series: Materials Science and Engineering; 2019: IOP Publishing.

27.   Shivakumar S, Mohana K. Studies on the inhibitive performance of Cinnamomum zeylanicum extracts on the corrosion of mild steel in hydrochloric acid and sulphuric acid media. Journal of Materials and Environmental Science. 2013; 4(3, Cop):448-59.

28.   Schorr M, Yahalom J. The significance of the energy of activation for the dissolution reaction of metal in acids. Corrosion Science. 1972; 12(11):867-8.

29.   Al-Saadie KAS, Al-Mashhdani HAY. Corrosion Protection Study for Caron Steel in Seawater by Coating with SiC and ZrO 2 Nanoparticles. American Journal of Chemistry. 2015; 5(1):28-39.

30.   Mohammed RA. An investigation of electropolymerization and corrosion protection properties of polypyrrole coating on carbon steel and stainless steel. Iraq: University of Baghdad; 2014.

 

 

 

 

Received on 14.10.2019           Modified on 24.12.2019